LOW RADIATION AND CONTRAST MEDIUM DOSE IN 64- SLICES MULTIDTECTOR CT ANGIOGRAPHY OF THORACIC AORTA
Ippolita Valentina Di Molfetta, Stefano Del Monte, Antonino Guerrisi, Giuseppe Guglielmo Aloise, Anna Forbidussi, Domenico Vito Di Molfetta, Basilio Lippi
Pages: 210-213
DOI: 10.21175/RadJ.2017.03.042
Received: 17 FEB 2017, Accepted: 27 MAY 2017, Published online: 23 DEC 2017
Abstract |
References |
Full Text (PDF)
The evaluation of chronic aortic diseases, many protocols of low radiation dose and low medium iodined contrast dose are performed.
The main aim of this study is to give a preliminary evaluation of dose reduction and iodined dose reduction. In our Hospital from February 2013 to November 2016 we selected 150 patients divided into two groups: 60 for our study and group of control of 90 cases.
All CT examinations were performed with a 64-MDCT scan. (Optima-CT GE Healtcare) Tube voltage was reduced in our study (80 kVp versus 120 in our standard) with automated current modulation system in both groups.
Concerning the iodined dose reduction, in the study groups it is strongly reduced (40 cc of 370 mg/ml versus 90 cc of 370 mg/ml): a mechanical power injector was used to administer contrast material via catheters (20-gauge) placed in antecubital vein at a flow rate of 4.5 ml/sec
Two radiologists qualitaively graded image quality of all cases defining the walls and enhancement of the lumen of the aorta. On the basis of criteria reported in the literature a five point subjective scale was used to grade image quality, from excellent (1) to non diagnostic quality(5). The reasons for degraded image quality were due to high BMI and consisted expecially in low signal/noise ratio and in two cases it was due to suboptimal contrast enhancement owing to poor bolus timing.
In the cases of low signal/noise ratio a smooth filter was applied to reduce the noise
The results of this study provide useful information about reduction of radation dose and medium iodined constrast.
Diagnostic quality of scan performed with low dose of iodine and radiation are overlays with the scans performed with standard protocol.
The study groups revealed a strong reduction dose in terms of DLP and quality of images was similar to the group of control.
- D. Marin et al.,“Low-Tube-Voltage, HighTube-current Multidetector Abdominal CT: Improved Image Quality and Decreased Radiation Dose with Adaptive Statistical Iterative Reconstruction Algorithm—Initial Clinical Experience,” Radiology, vol. 254, no. 1, pp. 145 – 153, Jan. 2010.
DOI: 10.1148/radiol.09090094 PMid: 20032149
- D. Ippolito et al., “Low kV settings CT angiography (CTA) with low dose contrast medium volume protocol in the assesment of thoracic and abdominal aorta disease: a feasibility study,” Br. J. Radiol., vol. 88, no. 1049, 20140140, May 2015.
DOI: 10.1259/bjr.20140140 PMCid: PMC4628465
- A. Seehofnerova at al., “Feasibility of low contrast media volume in CT angiography of the aorta,” Eur. J. Radiol. Open, vol. 2, pp. 58 – 65, Apr. 2015.
DOI: 10.1016/j.ejro.2015.03.001 PMid: 26937437 PMCid: PMC4750622
- E. Cakmackci et al., “CT-angiography protocol with low dose radiation and low volume contrast medium for non-cardiac chest pain,” Quant. Imaging Med. Surg., vol. 4, no. 5, pp. 307 – 312, 2014.
DOI: 10.1016/j.ejro.2015.03.001 PMid: 26937437 PMCid: PMC4750622
- M. Yamamuro et al., “Coronary Angiography by 64-Detector Row Computed Tomography Using Low Dose of Contrast Material with Saline Chaser: Influence of Total Injection Volume on Vessel Attenuation,” J. Comput. Assist. Tomogr., vol. 31, no. 2, pp. 272 – 280, Mar-Apr. 2007.
DOI: 10.1097/01.rct.0000236422.35761.a1 PMid: 17414766
- A. Winklehner et al., “Automated attenuation-based tube potential selection for Thoracoabdominal computed tomography angiography,” Invest. Radiol., vol. 46, no. 12, Dec. 2011.
DOI: 10.1097/RLI.0b013e3182266448 PMid: 21730872
- T. B. Kyongtae, “Intravenous cotrast medium administation and scan timing at CT: Consideratins and Approaches,” Radiology, vol. 256, no. 1, pp. 32 – 61, Jul. 2010.
DOI: 10.1148/radiol.10090908 PMid: 20574084
- Y. Nakayama et al., “Lower tube voltage reduces contrast material and radiation doses on 16-MDCT aortography,” AJR Am. J. Roentgenol., vol. 187, no. 5, pp. W490 – W497, Nov. 2006.
DOI: 10.2214/AJR.05.0471 PMid: 17056879
- Y. Shen et al., “High-Pitch, Low-Voltage and Low-Iodine-Concentration CT Angiography of Aorta: Assessment of Image Quality and Radiation Dose with Iterative Reconstruction,” PLoS ONE, vol. 10, no. 2, e0117469, Feb. 2015. DOI: 10.1371/journal.pone.0117469
PMid: 25643353 PMCid: PMC4314070
- M. M. Mourits et al., “Reducing contrast medium volume and tube voltage in CT angiography of pumonary artery,” Clin. Radiol., vol. 71, no. 6, p. 615, Jun. 2016.
DOI: 10.1016/j.crad.2016.03.005 PMid: 27059387
|